AquaMax
®
2000/4000 Microplate Washer User Guide
124
0112-0151 K
General Commands
Initialization
Reset
Command: /
RESET
[no arg]
Software Reset
Command: /
SOFTRES
[no arg]
Action:
Resets the washer. Stops pumps, closes all valves, homes all
axes, and moves the plate carrier to the loaded position. It
requires polling of the instrument status for completion with
“/status” command.
Response:
“No data OK” response if the command succeeds.
Action:
Restores the instrument to turned on state without
disrupting the current communications connection, it does
not home axes.
Response:
“No data OK” response if the command succeeds.