![Molecular Devices AquaMax 2000 User Manual Download Page 138](http://html1.mh-extra.com/html/molecular-devices/aquamax-2000/aquamax-2000_user-manual_1823779138.webp)
AquaMax
®
2000/4000 Microplate Washer User Guide
138
0112-0151 K
Function-specific Step Arguments and Return Data
The <
code
> and <
tokens
> parameters of the /
STEP
command and response
are mnemonics for protocol steps primitives and required parameters.
<Code> is one of the following step types (all are four characters except
nop):
<Tokens> is the step parameter list in the format of single letter tokens
followed by corresponding (integer) values, where the tokens are:
disp
dispense
aspi
aspirate
wash
wash
shak
shake
soak
soak (wait)
repe
repeat
prim
prime
purg
purge
stan
standby
rflu
rinse-flush
nop
blank (no operation)
whsq fast
wash
I
inlet number, 0-3 (0=green, 1=red, 2=blue, 3=yellow), -1
(prompt)
R
rate, generally 0-4
T
time, seconds
U
time, tenths of seconds
V
volume, microliters/well
Summary of Contents for AquaMax 2000
Page 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Page 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Page 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Page 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Page 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Page 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Page 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...