Operation: The Touch Screen Interface
0112-0151 K
51
Table 4-12:
Wash settings
Item
Description
1
Inlet–Liquid to be used for this step.
The available options correspond to the color-coded reagents
connected to the washer.
2
Prompt–Indicates that the system will pause before dispensing and
wait for the user to specify an inlet.
3
Rate–Wash rate (see
). This option is available
for the AquaMax
®
4000 Microplate Washer only.
4
Volume–Volume of wash fluid used per well. Use 50
μ
L to 3000
μ
L.
5
Wash position–Position of aspiration probes during the wash.
•
Bottom: The aspiration probes aspirate to the bottom of the well,
then wash with the specified amount of wash fluid at that
position. Bottom washing results in partially empty wells at the
end of the wash.
•
Top: The aspiration probes remain just below the top of the well,
then wash with the specified amount of wash fluid at that
position. Top washing results in full wells at the end of the wash.
Table 4-13:
Wash Rates
Model #
Options
96 (
μ
L/well/sec)
384 (
μ
L/well/sec)
AQ 4000
Fast
270
70
Gentle
200
60
1
4
3
5
2
Summary of Contents for AquaMax 2000
Page 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Page 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Page 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Page 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Page 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Page 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Page 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...