Robot Command Interface
0112-0151 K
143
Shake (SHAK)
The plate is shaken for the specified time at the given intensity level.
Shaking is linear only (along horizontal axis).
Parameter tokens are:
Soak / Wait (SOAK)
Needles are removed from the wells and no other operations are
performed for the specified time.
Parameter tokens are:
S
Speed: 0 =slow, 1=medium, 2=fast
T
Time: 1-600 [sec]
Example:
/step set 2 SHAK T120
CR
returns OK
TABCRLF
>
/step get 2
CR
returns OK
TAB
SHAK T120
CRLF
>
T
Time: 1-7200 [sec]
Example:
/step set 2 SOAK T60
CR
returns OK
TABCRLF
>
/step get 2
CR
returns OK
TAB
SOAK T60
CRLF
>