![Mohr CT100B Series Operator'S Manual Download Page 42](http://html.mh-extra.com/html/mohr/ct100b-series/ct100b-series_operators-manual_1823448042.webp)
3. Operating Instructions
Table 3.4 (Continued) MENU Button and Menus
Menu Selection
Purpose/Action
Set
Ω
/(unit) Pos.
Set the position of the ohms/(unit distance) position
to that of the active cursor.
Clear
Reset all Ω/(unit) settings.
Math
Settings
Meas.
Settings
See Meas. Settings under Measurement menu above.
Network Settings
Static Network Settings
Modify static network settings (if DHCP inactive):
IP Address
. . . . . . . . . . . . . . . . . . . . .
IP address (xxx.xxx.xxx.xxx)
Netmask
. . . . . . . . . . . . . . . . . . . . . . . .
Subnet for CT100B
Gateway (optional)
. . . . . . . . . . .
Gateway server setting
Nameserver 1 and 2
(optional)
. . . . . . . . . . . . . . . . . . . . . . . . . .
Nameserver settings
DHCP on/off
Toggles DHCP on/off.
Show IP Config.
Displays current Ethernet configuration.
Web Export On/Fast/Off
Activates and toggles function of web server into
three states: On, Fast Mode, and Off. The web
server provides .csv format files as well as screen
capture bitmaps. See Section 3.11.25 for more
details on accessing data using web export.
On.............................
Normal operation (~2 frames/sec).
Fast Mode.....................
Fast refresh (~7 frames/sec).
Off............................
Web server not active.
[
Note:
See Section 4 for more information on using
CT Viewer 2 to acquire TDR traces. See
for information on the CT100B
Python programming library and binary programming
protocol for Ethernet CT100B interface.]
Selected .csv Options
Opens a dialog for adding additional data columns
in the .csv trace export file.
Set Listen Ports
This dialog box allows you to define a different RPC
port # and a different Remote Control Port # for
use with the CT Viewer 2 application (default 12347)
SP232 Protocol On / Off
Toggles SP232 emulation on/off. A USB to serial
adapter is required for SP232 emulation. Press this
button while emulation is on to switch the baud rate.
Power
DC Power Save
Sets the inactivity timeout for power save mode
while the device is running on external DC supply.
DC Shutdowns
Sets the inactivity timeout for shutdown while the
device is running on external DC supply.
24