B. Operating Performance Checks
B.1. General Information
The CT100B Operator Performance Checks are a series of procedures used to verify the
calibration of a CT100B TDR. The following series of checks should be performed after a unit
has been newly calibrated to verify compliance to published specifications, but may also be
performed to determine if calibration is required. It is recommended to perform these checks at
least every 6 months. See Section D.4 for more information on calibration.
The following series of steps are designed to allow a CT100B operator to completely verify the
properties of an individual CT100B unit to each published specification. Throughout this
procedure any references made to the CT100B should be considered applicable to the CT100HF
and CT100S.
NOTE: Operator Performance Checks should not be considered CT100B
calibration.
NOTE: If a CT100B fails any Operator Performance Check, it should be
serviced by a qualified repair facility.
B.2. Required Equipment
Table B.1.
Required Performance Check Equipment for CT100B
Item
MOHR part number
Precision 50 Ω terminator
CT100-AC-ER50-S
Connector, SMA female to BNC male
CT100-AC-ISFBM
50 Ω 36 in. reference cable
CT100-AC-W536
Shorting cap
CT100-AC-IBS
BNC male to male adapter
CT100-AC-IBMM
Table B.2.
Required Performance Check Equipment for CT100HF and CT100B with
SMA Option
Item
MOHR part number
Precision 50 Ω terminator
CT100-AC-ER50-S
CT100B TDR Cable Analyzers Operator’s Manual