![Mohr CT100B Series Operator'S Manual Download Page 102](http://html.mh-extra.com/html/mohr/ct100b-series/ct100b-series_operators-manual_1823448102.webp)
A. Specifications
Characteristic
Specifications
Notes
Horizontal position
Any distance to full scale
can be moved on the
screen.
Vp range
0.250000 to 1.000000
Vp default resolution
0.001
Vp fine resolution
0.000001
Vp accuracy (coarse)
±
0
.
1%
Vp accuracy (fine)
±
0
.
0001%
USB host port
USB 1.1
Front USB host connector
is powered (500 mA limit)
with 500 VDC isolation
from test port.
USB client port
USB 1.1
500 VDC isolation from
test port.
Ethernet port
The instrument includes a
10/100mbps Ethernet
port for network
communication.
500 VDC isolation from
test port.
DC power supply
24 VDC, 2.5 A, positive
tip. Use only the
MOHR-approved power
supply.
500 VDC isolation from
test port.
Battery pack
12 NiMH AA cells, fused
at 2.5 A
Pack may wear out over
time.
Battery operation time
>
6 hours. Power save
options may extend
operating time.
18V full charge down to
12V battery low, 35%
duty cycle.
Battery charge time
2.5 hours from fully
discharged state
Overcharge protection
Charging discontinues
once full charge is
attained.
Discharge protection
Instrument turns off prior
to battery damage
Software shutdown when
battery is low.
Charge capacity
2.7 Amp-hours
(nominal-new)
A.2. Environmental Specifications
A.2.1
Temperature
Operating temperature
0
◦
to +50
◦
C
Non-operating temperature
-20
◦
C to +60
◦
C
A.3. Mechanical Specifications
A.3.1
Weight
Without cover
2150 g (4.740 lb.)
With cover
2323 g (5.121 lb.)
84