B. Operating Performance Checks
B.4.18
Horizontal Scale Check
1. From the
Main
menu, set
Cable Len
to “Long”.
2. Using the
M-FUNC
button and
M-FUNCTION
knob, set Vp to 1.00.
3. Turn the
HORIZONTAL SCALE
knob counter clockwise. Horizontal scale should increase
beyond 61 m/div or 200 ft./div.
4. Turn the
HORIZONTAL SCALE
knob clockwise. Horizontal scale should decrease below
0.01 m/div or 0.03 ft./div.
B.4.19
Horizontal Range Check
See the cursor readout range instructions on page 90. Horizontal range and cursor readout range
are always equal.
B.4.20
Horizontal Position Check
This test is passed implicitly when the Cursor Readout Range and Horizontal Range checks are
passed.
92
CT100B TDR Cable Analyzers Operator’s Manual