A. Specifications
A.1. Electrical Specifications
Characteristic
Specifications
Notes
Reflected rise time, CT100B 10-90%
150 ps typ, 200 ps max
10 to 90%, into 50Ω
Reflected rise time, CT100B 20-80%
100 ps typ, 150 ps max
20 to 80%, into 50Ω
Reflected rise time, CT100HF 10-90%
100 ps typ, 130 ps max
10 to 90%, into 50Ω
Reflected rise time, CT100HF 20-80%
60 ps typ, 90 ps max
20 to 80%, into 50Ω
Jitter
15 ps max (peak-to-peak)
< 1 ps rms typ.
Output impedance
50Ω
±
2%
Pulse amplitude
300
±
10 mV
Averaged value into 50Ω
Pulse width
Fixed: 1.000-2.000,
5.000-6.000,
20
.
00
−
21
.
00
µ
s
Averaged values change
for short, medium, and
long cable length settings,
respectively
Pulse repetition time
Fixed: 4
.
000
±
0
.
002,
20
.
00
±
0
.
02,
80
.
00
±
0
.
02
µ
s
Averaged values change
for short, medium, and
long cable length settings,
respectively
Sequential sampling rate
Fixed: 250
.
0
±
0
.
2,
50
.
00
±
0
.
04,
12
.
50
±
0
.
02 kHz
Averaged values change
for short, medium, and
long cable length settings,
respectively
Vertical scale
0.5 m
ρ
/div
Vertical accuracy
±
3% full scale
<
±
1% of measurement
typ.
Vertical position
Any trace point is
movable to the center of
the screen.
Sampling efficiency, CT100B
50 to 90%
into 50Ω
Sampling efficiency, CT100HF
45 to 90%
into 50Ω
Noise
±
5 m
ρ
peak-to-peak
Input susceptibility
±
1 A max
into diode clamps
Distance cursor resolution
1/45 of one major division
Cursor readout range
≤
-0.3 m (-1 ft.) to
>
10000 m (33000 ft.)
Cursor readout resolution
0.0002 m (0.0007 ft.)
Distance measurement accuracy
<
1 cm (0.3 in.) or
<
0
.
5% of measurement
plus uncertainty in Vp,
whichever is greater
<
±
0
.
05% of
measurement typ.
Impedance readout range
<
1Ω to
>
1
.
5kΩ
Impedance resolution
3 significant digits
Impedance accuracy
±
10%, relative
measurement
±
2%
<
±
1%typ.(0 to 1kΩ)
Horizontal scale
0.01 m/div (0.03 ft/div)
to
≥
61 m/div (200
ft/div)
(variable with pulse
width)
Horizontal range
<
0 m to
>
1000 m
(3300+ ft)
CT100B TDR Cable Analyzers Operator’s Manual
83