3. Operating Instructions
Table 3.4 (Continued) MENU Button and Menus
Menu Selection
Purpose/Action
Reset
Deletes/restarts captured TDR trace envelope data.
Fill Mode Off/On
Toggles between fill and probability density plot
mode. Use fill mode to find magnitude of impedance
change and highlighting fault location. Use
probability density plot mode to find infrequent
transient faults.
Hold Off/On
When Hold is on, trace controls are locked for an
envelope plot. When Hold is off, trace controls work
normally.
Clear Historic
As trace controls are used to change an envelope
plot, the envelope plot preserves a record of the trace
from previous positions. Clear Historic erases this
history, leaving only the plot from the current
position.
(Horizontal Units)
Toggles through horizontal units (meters, inches,
feet, yards, centimeters)
Meas.
Settings
Horiz.
Units
Toggles through horizontal units (meters, inches,
feet, yards, centimeters)
RRC Classic/1502C Mode
Classic: Referenced to amplitude at inactive cursor.
1502C: Referenced to amplitude at test port.
Vp 3/6 Sig Figs
Toggle Vp significant digits from 3 to 6 digits.
Vertical Correction
Adjust m
ρ
Offset
Allows manual vertical measurement offset
adjustments. This allows advanced users to
manually calibrate impedance measurements at a
particular location to a known value.
When activated the
M-FUNCTION
knob adjusts
the vertical offset. The blue
M-FUNC
indicator
shows the offset in m
ρ
(millirho).
Cancel m
ρ
Offset
Use this option to cancel a manual m
ρ
(millirho)
correction. Automatic temperature adjustment will
resume auto-correction of the vertical measurements.
Adjust
Ω
/(unit)
Use the
M-FUNCTION
knob to change the ohms
per unit distance resistive loss setting until the cable
shows no upward or downward trend across its
length.
Ω
/(unit) Correct.
On/Off
Toggle resistive cable loss correction on/off, as
described above. If the end point of the correction
has not been set, then it will be set to the position of
the active cursor.
CT100B TDR Cable Analyzers Operator’s Manual