C. Operator Troubleshooting
C.3.7
TDR Trace Verification
Return to Main Menu and navigate
to Settings
à
Diagnostics and select
Analog. Attach a shorted
terminator and press OK
Did all tests
displayed pass?
NO
FAILED
YES
Using an oscilloscope,
verify the TDR pulse
characteristics as
described in “Operator
Performance Checks”
Are the pulse
characteristics within
the device’s
specifications?
NO
YES
Start TDR Pulse
Diagnostics
Attach a 50 ohm test
cable of known
length and set Vp to
match the pulse
velocity of the cable
Measure the length
of the cable for
Short, Medium, and
Long Cable Length
Settings
Do distance
measurements meet
specification?
PASS
TDR Trace
Verification Passed
Enable temperature
display. This can be
found under the
Main Menu
à
Settings
à
Display
Do temperatures update
on-screen, and within ± 15ᵒ of
ambient?
NO
YES
FAIL
FAIL
Verify the Vertical
Accuracy per the
“Operator Performance
Checks”
PASS
100
CT100B TDR Cable Analyzers Operator’s Manual