Instructions for use
LSI_K_np_en 1206_mmm_V2.08_B2V
35
c) Ventilator speed cannot be set in SS
programme; it is always maximum (100 %).
d) In the two-door model, if no printer is
connected or if a printer is connected but the
operator does not want to print, the value
“Printer switched off” must be set in the Print
mode of Service 04, otherwise the display on
the unloading (clean) side of the device does
not work.
11.16.2
STERILIzATION
PROGRAMME
SP
WITH
PREHEATING
PHASE
This device setting is used if we want the whole
load to be heated evenly already at the beginning
of the sterilization as such. The required
temperature, and especially the preheating
duration must be determined experimentally,
or the operator’s experience must be used for
different materials and their quantity in the load.
The programme works with two segments on 2
temperature levels: T1(preheating time) and T2
(sterilization/exposure time). The duration of T1
level is limited while the duration of T2 level can
be either limited or unlimited. T1 may be higher
or lower than T2 (depends on the operator’s
requirements or experience).
Ventilator speed cannot be set in SP programme;
it is always maximum (100 %).
11.16.2.1
SP
PROGRAMME
SETTING
11.16.2.2
SEGMENT
SETTING
FOR
T
1
T1 temperature setting:
move the cursor to the Programme position
and switch to SP programme. Place the cursor
into the graph by means of ◄► and check
whether T1 ( /¯\_ ) is activated (the cursor is on
T1 segment). If it is not activated, activate it by
means of ▲.
Go to position °C – temperature setting – and
set the temperature.
Temperature T1 delay setting:
move the cursor below the eleventh character
of the display and recall the information about
parameters by means of ▲ pressing. The display
shows “t” (see the notes above). Enter the
position and set the delay time between 00 h
00 min and 99 h 59 min.
Temperature T1 and delay displaying
11.16.2.3
SEGMENT
SETTING
FOR
T
2
T2 temperature setting:
Place the cursor into the graph and check
whether T2 ( /¯\_ ) is activated. If the segment is
not active, switch the cursor position by means
of ▲. Move the cursor to °C position and set the
temperature.
Temperature T2 delay setting:
move the cursor below the eleventh character
of the display and call the information about
parameters by means of ▲ pressing. The display
shows “t”. Enter the position
and set the delay
time between 00 h 00 min and 99 h 59 min, or
unlimited ∞:∞ (set the infinity by setting 99 hours,
moving the cursor to the figure 9 in units and
pressing ▲).
Temperature T2 and delay displaying:
11.16.2.4
SP
PROGRAMME
START
Press
STArT/STOp
. you will hear a sound
signal and the propeller symbol in the first
character starts rotating clockwise. See the
display below:
The graph shows the actual programme status.
Program Segment
°C
RH
5 6
S P /
O
O
O .
1
0
0
O : O
/ –
=
t O
O
–
–
–
–
–
– –
–
–
Program Segment
°C
RH
5 6
S P /
O
O
O .
1
0
0
O : O
/
–
=
t O
O
–
SP=
T1 T2
S1 S2 S3 S4