The first segment in the graph is active
°C°C°C = actual temperature
hh:mm = programmed time of delay, time
countdown to zero.
The second segment in the graph is active;
the first segment has elapsed
°C°C°C = actual temperature
hh:mm = time from the start till the preset
temperature reaching.
The third segment in the graph is active
The preset temperature has been reached. If the
procedure duration is unlimited, elapsed time is
displayed. If the procedure duration is limited,
time remaining till its end is displayed. after the
programme ends, chamber temperature and the
text “End” are displayed. Both the indication of
the programme and the chamber temperature
disappear after pressing any key on the panel.
11.16.1.4
INFORMATION
ON
SET
PARAMETERS
DURING
OPERATION
OF
THE
UNIT
See item 4.2.4.
Instructions for use
34
LSI_K_np_en 1206_mmm_V2.08_B2V
=
/
–
= t
ř
etí segment
segment 3
11.16.1.5
STOP
SS
Press
STArT/STOp
.
11.16.1.6
STERILIzATION
PROGRAMME
SETTING
AND
COURSE
If the ventilation flap or door are opened
during the delayed start phase and rise to the
sterilization temperature, message “Open” is
displayed on the right bottom side. The device
continues working without interruption.
Exposure time countdown starts several seconds
after the sterilization temperature was reached.
If the door or the ventilation flap are opened in
this phase, the display shows “Open”. If you
do not close the flap or the door, the device is
switched off after the exposure time ends. The
display shows “Open”.
If you close the door or the flap during the
exposure, the exposure (or temperature rise
according to the actual temperature in the
chamber) starts again from the beginning.
If the temperature drops below the preset one in
this phase:
a) If caused by a power supply failure, a new
cycle is started from the beginning of the
temperature rise after the power supply
recovery;
b) If caused by any other reason, the display
shows Error 02.
Dry-heat sterilization parameters
:
Temperature [°C]
Sterilization exposure [min]
160
60
170
30
180
20
Note:
a) In sterilization programmes, the temperature
deviation in the sterilization space within
the range of sterilization temperatures
(definition see in Chapter Working Space) is,
in accordance with the national regulations,
from -1 °C to +5 °C as compared to the
preset value. If Service 48 is activated, the
display shows Error 02 and Error 03 if the
temperature drops below the given limit and
exceeds the given limit, respectively.
b) In special cases, the sterilization cycle can
be started after the required temperature is
reached on the independent flexible sensor
PT2 placed by the user so that it could read
the temperature of the sterilized material. The
required device resetting may be carried out
by the service technician.
Program Segment
°C
RH
5 6
* S
/
–
°C
°C
°C .
1
0
0
h
h
: min
=
Program Segment
°C
RH
5 6
* S
// – °C
°C
°C .
1
0
0
h
h
: min
–
//–
= druhý segment
–
segment 2
Program Segment
°C
RH
5 6
* S
/ – °C
°C
°C .
1
0
0
h
h
: min
=
=/
–
= první segment
segment 1