Warning:
The plug must be connected only by a person
with a sufficient qualification.
When the plug is plugged in and out of the
socket, or when manipulated with the inner
equipment, must INCUCELL V be switched off
(the mains switch in the position-„0“).
The socket-plug contact must be tightened in
order to prevent disfunctioning of the protective
system.
After disconnecting the socket-plug contact fit
carefully the protection covers on the socket and
the plug so that impurities and humidity would not
get inside.
11.10
DEVICES
WITH
THE
DOOR
bLOCKING
FUNCTION
In this model, electrically controlled mechanism
is used to prevent unwanted door opening. The
mechanism remains in its position even after the
device is switched off.
ATTENTiON!
Blocked door may not be opened forcibly. It may
damage the blocking system mechanically. If
you need to open the door in case of emergency
(e.g. power supply failure), follow the chapter
Emergency Door opening.
11.10.1
ONE
-
DOOR
MODELS
Two ways of blocking can be selected in
SErVICE 58:
1. manual - The door is blocked/unblocked by
pressing ◄ and ► buttons for 2 seconds.
If you try to block an open door, you are
warned of an error by a prolonged beep.
2. automatic - Close the door. It is blocked
automatically after preset time elapsing.
Unblock the door by pressing ◄ and ►
buttons for 2 seconds. The door is blocked
only when the programme is switched on.
When you unblock the door, you have a preset
period of time to open it. If you do not open it,
the door will be reblocked after the preset time
elapsing (2 to 9 s).
The blocked door is indicated on the display by
the flashing character “#”.
The door unblocking can be conditioned by the
So card insertion.
Instructions for use
30
LSI_K_np_en 1206_mmm_V2.08_B2V
11.10.2
TWO
-
DOOR
VENTICELL
The blocking function is pre-set in Service 58.
The setting is then applied to all programmes.
If “B=Lck” is set for “End”, “End” will be changed
automatically to “Stp” after reaching the given
state.
you can use the button on the unloading side
panel for switching from “Stp” to “End” (out of
function when any door is open) and from “End”
to “Stp” (out of function when the unloading door
is open).
In the sterilization programmes (see chapter
Venticell with the sterilization function), you can
change from “End” to “Stp” by the said button
only if the unloading door was opened before
(material unloading).
11.11
PROTECTION
OF
PROGRAMME
SETTING
bY
MEANS
OF
A
CHIP
CARD
Selected programme parameters may only be
changed after So card insertion. If the card is not
inserted, the programmes are displayed in small
letters (p1 to p6). after the card is inserted, they
are displayed in capital letters (P1 to P6).
11.12
DOOR
OPENING
SIGNALLING
Door opening during the programme run is
signalled both on the display and by a sound
signal. Sound signalling can be temporarily
switched off by any arrow pressing.
11.13
TWO
-
DOOR
PASS
-
THROUGH
MODEL
available in VC 222, 404 and 707 only. allows
material loading in one area and its unloading in
another area after heat treatment.
11.13.1
VENTICELL
222/2
-
INSTALLATION
DATA
1) Input power 3700 W
2) Electrical installation – connection of the unit:
– plug VDE 0623, DIN 4962/63in, CEE 17,
IEC 309, 3P + N + PE, 16a/380-415 VaC
– supply to the socket from the switchboard
5x2.5 mm
2
Cu, protection by circuit breaker
16 a – place near the unit
– cable length 3 m
3) Heat emission at 200 °C: 950 W
4) Weight: 105 kg
Basic dimensional data are the same as
In case of the following VC 404 / 2D – with
the exception of the height – see the table of