4.1.10
SERVICE
58
-
DOOR
bLOCKING
This service is available only in models equipped
with the door blocking function.
One-door model:
Press
STArT/STOp
and open the service.
Select one of the following possibilities by means
of ▲:
1. Inactive – the door is unblocked permanently
2. manual – the door can be blocked and
unblocked by means of ▲ and ►
3. automatic – the door is blocked automatically
after a pre-set period of time elapses after the
door closing.
The time can be set within the range from 2 to 9
seconds. Unblock the door by pressing ▲ and ►
buttons for 2 seconds.
You can select “C” on the last display position. In
such case, the door can be unblocked only after
inserting the So card.
S e r v i c e
5 8
A u t o m . a f t e r
2 s
C
Two-door VENTiCELL models:
after the service is started, the cursor is placed
on the first left position; four statuses can be pre-
set on the bottom line by means of ▲:
S
t r : A
=
O p n
B
=
L c k
This status defines the door blocking after the
programme start by means of
STArT/STOp
.
S
t p : A
=
O p n
B
=
L c k
This status defines the door blocking after the
programme interruption by means of
STArT/
STOp
.
E
n d : A
=
L c k
B
=
O p n
This status defines the door blocking after the
programme termination (the panels show “Konec
Px” and “End”.
U
n l o c k
a l
l ( X / W )
Press
X/W
to unblock both doors.
Use ◄ and ► to move the cursor between “A” =
loading door, and “B” = unloading door.
Three values are defined for these options (the
tables show the initial setting):
•
„Opn“ – unblocked
•
„Lck“ – blocked
•
„Txyz“ – the door will be unblocked after the
temperature in the chamber drops below
“xyz”. “xyz” can be pre-set by means of ▲ in
the range from 50 to 200 °C.
It is recommended to pre-set the same “xyz”
value as in Service 06 (duration of the ventilator
run after the programme termination).
“A=Lck” cannot be pre-set if the device is in “Stp”
status. “A=Lck” only can be pre-set if the device
is in “End” status.
The device will be in “Stp” status only if the
programme run is interrupted by means of
STArT/STOp
; however, this shall not be applied
if an error status is cancelled by means of
the said button. The door blocking will remain
unchanged after the error status cancellation.
See more in the Chapter optional accessories.
4.1.11
SERVICE
62 –
SOUNDS
SETTING
This service allows individual switch-on and -off
of the following events:
- Device switch-on;
- Device errors;
- operator’s errors;
- Buttons pressing;
- Programme start;
- Programme end;
- Door opening during the programme run.
4.1.12
SERVICE
64 –
PROGRAMME
P
6
SHIFTING
This service shifts the programme start to a new
time. See more in chapter Programme P6.
4.1.13
SERVICE
09
-
CANCEL
THE
SERVICE
MODE
By pressing
STArT/STOp
, the service mode is
cancelled. you can then modify programmes and
start them freely.
4.2
PROGRAM
P1
Constant temperature is required.
P 1
=
The cabinet operates continuously at set
temperature until switching the unit off.
4.2.1
SET
P
1
Put cursor to the position Program with help of
◄► and switch to P1 with help of ▲. Set the
temperature at the position of °C within the range
10
LSI_K_np_en 1206_mmm_V2.08_B2V
Instructions for use