Appendix B: Valve Orifice Selection
How To Verify the Orifice Selection
59
3.
Use the index number and your maximum flow rate, to determine the orifice size from
Figure 24.
Each line represents the maximum flow rate for the orifice. Choose the orifice number
above your point on the graph to ensure that the orifice can deliver the required flow.
Continuing with the example above, the index number is 175, and assuming a maximum
required flow rate of 1000 sccm, the correct orifice would be number 2.
Note
If the point on the graph falls close to the maximum flow rate for an
orifice, you may choose to use the next largest orifice number. However,
a larger orifice number provides a higher minimum flow rate as well. For
example, in an elastomer sealed unit, orifice number 2 covers the range
from 1 to 1000 sccm (N
2
at 1 atm
∆
P), whereas orifice number 3 covers
from 5 to 5000 sccm.
Index Number
M
a
xi
m
u
m
F
lo
w
(s
ccm
)
Ma
x
im
u
m Flo
w
(
s
lm
)
1
10
100
1000
0.001
0.01
0.1
1
10
100
1000
1
10
100
1000
10000
100000
1000000
3
2
1
A
Figure 24: Flow Range Selection
4.
Check the orifice size of your 640 controller (included in the model number).
Summary of Contents for 640A/641A
Page 4: ...116669 P1 Rev C 12 99 MKS Type 640A 641A Pressure Controller...
Page 9: ...Table of Contents vi...
Page 11: ...List of Figures viii...
Page 13: ...List of Tables x...
Page 21: ...Customer Support Chapter One General Information 8 This page intentionally left blank...
Page 45: ...Initial Configuration Chapter Two Installation 32 This page intentionally left blank...
Page 85: ...Model Code Description Appendix D Model Code Explanation 72 This page intentionally left blank...