Chapter Four: Operation
How To Check the Zero
43
Chapter Four: Operation
How To Check the Zero
Check the transducer zero before operating the unit initially and then periodically as required.
The zero can be set (or reset) by adjusting the zero potentiometer located on the top cover of the
transducer or, on the front panel of an MKS Power Supply/Readout, if you are using one.
Note
The port labeled “Controlled Pressure Port” serves as the transducer inlet
port. Figure 3, page 17, shows the labels on the unit.
How To Zero a Type 640 (Absolute) Unit
To zero a 640 controller (with an absolute transducer inside) you must pump the unit, with the
power on, down to a pressure less than the transducer’s resolution (0.01% of Full Scale).
Note
The zero adjustment must be made at a pressure less than the transducer’s
resolution (0.01% of F.S.).
In addition, you should position the unit in the same orientation as it will
be positioned when installed in your system.
Zeroing a transducer above its stated minimum resolution creates a zero offset relative to true
absolute pressure. All subsequent readings are then linear and accurate relative to the offset
value.
Note
If your system cannot achieve a sufficiently low pressure to set the
transducer zero, you may use a vacuum leak detector with sufficient
vacuum pumping (to achieve a proper zeroing pressure). In this case,
mount the unit on the leak detector in the same plane of orientation as it
will be during actual use.
Summary of Contents for 640A/641A
Page 4: ...116669 P1 Rev C 12 99 MKS Type 640A 641A Pressure Controller...
Page 9: ...Table of Contents vi...
Page 11: ...List of Figures viii...
Page 13: ...List of Tables x...
Page 21: ...Customer Support Chapter One General Information 8 This page intentionally left blank...
Page 45: ...Initial Configuration Chapter Two Installation 32 This page intentionally left blank...
Page 85: ...Model Code Description Appendix D Model Code Explanation 72 This page intentionally left blank...