Chapter 6: Operation (Profibus Units)
1579A/579A
50
Small Setup
Name: SMALL_SETUP
Type:
Initial Write (small)
Size: 19
Parameters:
Name
Add.
Type
Comment
STRUCT_ID
0:0
uint8
0x10 (SMALL_SETUP)
INITIAL_SETUP
2:0
uint:1
THIS, ROM
BASE_UNIT
2:1
uint:1
Display in base unit
OPERATION_MODE
2:2
uint:1
0=ANALOG, 1=PROFIBUS
Reserved
2:3
uint:5
Reserved
1:0
uint:8
USER_SPAN
3:0
long
5% .. 200% in [%] (in 1E-4 steps)
USER_ZERO
7:0
long
-5% .. +5% of full scale (in 1E-4 steps)
HIGH_TRIP_POINT
11:0
long
-10% .. +120% of full scale (in 1E-4 steps)
LOW_TRIP_POINT
15:0
long
-10% .. +120% of full scale (in 1E-4 steps)
Summary of Contents for 1579A
Page 8: ...Table of Contents 1579A 579A viii This page intentionally left blank...
Page 12: ...Safety Information 1579A 579A 12 This page intentionally left blank...
Page 16: ...Chapter 1 General Information 1579A 579A 16 This page intentionally left blank...
Page 20: ...Chapter 2 Shipment 1579A 579A 20 This page intentionally left blank...
Page 28: ...Chapter 3 Connectors and Control Elements 1579A 579A 28 This page intentionally left blank...
Page 38: ...Chapter 5 Operation Analog Units 1579A 579A 38 This page intentionally left blank...
Page 70: ...Chapter 10 Maintenance 1579A 579A 70 This page intentionally left blank...
Page 74: ...Appendix A Product Specifications 1579A 579A 74 This page intentionally left blank...
Page 82: ...Appendix C Gas Correction Factors 1579A 579A 82 This page intentionally left blank...
Page 84: ...Appendix C Gas Correction Factors 1579A 579A 84 This page intentionally left blank...