1579A/579A
Chapter 6: Operation (Profibus Units)
55
Name
Add.
Type
Comment
STANDARD_TEMP 114:0
long
0 °C (in 10E-4 steps)
STANDARD_PRESSURE 118:0 long
101.1kPa (in 10E-4 steps)
VALVE_TYPE 122:0
uint8
0=SOLENOID, 1=VOICE_COIL,
2=PIEZO_ELECTRIC
VALVE_POWER_OFF_
MODE
123:0 uint8
0=CLOSED, 1=OPEN, 2=LAST_POS
GAS_TABLE_NUM
124:0
uint8 Number of gas tables programmed.
i.e. values != 0 in GAS_CODE_OF_TABLE_I
GAS_CODE_OF_
TABLE_I
125:0
uint8[15]
gas code of gas tables
0=no table (y=x)
POINT_NUM_OF_
TABLE_I
140:0 uint8[15] Point number of gas tables
TABLE_FLAGS 155:0
uint:1[15] 0=FACTORY; 1=USER; 1 means that the user
has overwritten the gas table once. All these flags
will be reset at factory setup.
ACTIVE_GAS_NAME 157:0
char[16]
e.g.
N2, is name of DEFAULT_TABLE if there
was no cyclic comm. In the past.
CAL_DATE 173:0
char[6]
MM/DD/YY
USER_TAG
579:0 char[32]
any 32 character string
REM_TIME_TO_CAL 211:0
uint16 Remaining
TIME_TO_CAL
FLOW_TOTALIZED 213:0
long in sl/sm3 (in 10E-4 steps)
i.e. min. 298 days for a 500 range.
No base unit feature.
Summary of Contents for 1579A
Page 8: ...Table of Contents 1579A 579A viii This page intentionally left blank...
Page 12: ...Safety Information 1579A 579A 12 This page intentionally left blank...
Page 16: ...Chapter 1 General Information 1579A 579A 16 This page intentionally left blank...
Page 20: ...Chapter 2 Shipment 1579A 579A 20 This page intentionally left blank...
Page 28: ...Chapter 3 Connectors and Control Elements 1579A 579A 28 This page intentionally left blank...
Page 38: ...Chapter 5 Operation Analog Units 1579A 579A 38 This page intentionally left blank...
Page 70: ...Chapter 10 Maintenance 1579A 579A 70 This page intentionally left blank...
Page 74: ...Appendix A Product Specifications 1579A 579A 74 This page intentionally left blank...
Page 82: ...Appendix C Gas Correction Factors 1579A 579A 82 This page intentionally left blank...
Page 84: ...Appendix C Gas Correction Factors 1579A 579A 84 This page intentionally left blank...