1579A/579A
Chapter 6: Operation (Profibus Units)
49
Full Receive Data
Name: FLOW_MFC
Type:
Cyclic Read (full)
Size: 18
Description:
Analog Input Transducer Block. This block is used if a full setup was
performed.
Parameters:
Name
Add.
Type
Comment
HIGH_LIMIT_ALARM
1:0
uint:1 (flow
> HIGH_LIMIT), Hysteresis = 0.5% *
LOW_LIMIT_ALARM
1:1
uint:1
(flow < LOW_LIMIT) , Hysteresis = 0.5% *
SYSTEM_ERROR
1:2
uint:1
any severe system error condition
HIGH2_LIMIT_ALARM
1:3 uint:1 (flow
> HIGH2_LIMIT), Hysteresis = 0.25%
LOW2_LIMIT_ALARM
1:4
uint:1 (flow
< LOW2_LIMIT), Hysteresis = 0.25%
VALVE_CLOSED
1:5
uint:1
(THERMAL_MASS_FLOW_RATE < 1%) &&
(VALVE_OVERRIDE == FLOW_OFF)
PURGE
1:6
uint:1
THERMAL_MASS_FLOW_RATE > 110%
OVER_TEMPERATURE
1:7
uint:1
INTERNAL_TEMP > MAX_TEMP
VALVE_DRIVE_ALARM
0:0 uint:1
VALVE_DRIVE_LEVEL > MAX_VTP
CALIBRATION_
RECOMMENED
0:1
uint:1
TIME_TO_CAL count down expired
UNCALIBRATED
0:2
uint:1
if a disabled or no table is used
CONTROLLER_ERROR
0:3
uint:1
abs (setp - flow) greater for a longer time period
MEMORY_FAILURE
0:4
uint:1
E2PROM checksum error
UNEXPECTED_
CONDITION
0:5
uint:1
any process error condition
Reserved
0:6
uint:2
THERMAL_MASS_FLOW
_RATE
2:0
long
in [FLOW_UNIT] in 10E-4 steps
INTERNAL_TEMP
6:0
long
temperature in [° C] (in 10E-4 steps)
VALVE_DRIVE_LEVEL
10:0
long
0 .. 100% (in 10E-4 steps)
0% = valve is closed
100% = valve is in purge position (full open)
FLOW_TOTALIZED
14:0
long
in sl / sm3 (in 10E-4 steps)
i.e. min. 298 days for a 500 range.
*) Hysteresis is +/- 0.25% (i.e. 0.5% in total) based on actual limit
Summary of Contents for 1579A
Page 8: ...Table of Contents 1579A 579A viii This page intentionally left blank...
Page 12: ...Safety Information 1579A 579A 12 This page intentionally left blank...
Page 16: ...Chapter 1 General Information 1579A 579A 16 This page intentionally left blank...
Page 20: ...Chapter 2 Shipment 1579A 579A 20 This page intentionally left blank...
Page 28: ...Chapter 3 Connectors and Control Elements 1579A 579A 28 This page intentionally left blank...
Page 38: ...Chapter 5 Operation Analog Units 1579A 579A 38 This page intentionally left blank...
Page 70: ...Chapter 10 Maintenance 1579A 579A 70 This page intentionally left blank...
Page 74: ...Appendix A Product Specifications 1579A 579A 74 This page intentionally left blank...
Page 82: ...Appendix C Gas Correction Factors 1579A 579A 82 This page intentionally left blank...
Page 84: ...Appendix C Gas Correction Factors 1579A 579A 84 This page intentionally left blank...