Defibrillator Proof, Type BF equipment with isolated (F) applied part. Each of the electrode
terminals of the Generator can withstand the effects of defibrillator discharge.
Liquid Spillage as per IEC 60601-2-2. The Generator enclosure will prevent reasonable
amounts of liquid from interfering with the Generator’s safe and satisfactory operation.
Intermittent Operation
The Generator is cooled by natural convection. Under maximum power setting and rated load
conditions the Generator is suitable for a 10 seconds on, 30 seconds off duty cycle for 1 hour.
Temperature Control Mode
Control range
45 – 95ºC
Display range
10 – 99ºC
Over temp indicator
>8ºC above SET temperature
VaPR footSWItCH SPeCIfICatIoNS
Cat.
No
. 225003 & 225023
Rating IPX8
OUTPUT WAVEFORM AND CHARACTERISTICS
(VAPR II ONLY)
Waveform
The RF output is a variable amplitude sinusoid
waveform varying between approximately
340kHz and 450kHz, corresponding to minimum
and maximum load impedance respectively.
Crest Factor
V1, V2, V3, BV1, BV2 and Desiccate
A nominal crest factor of 1.4 for all outputs.
Power
Maximum power 200 watts into 160 ohms.
Max Voltage:
V1
254V RMS
V2
307V RMS
V3
340V RMS
BV1
307V RMS
BV2
340V RMS
Desiccate
120V RMS
VAPR TC Desiccate
100V RMS
NOTE
The following load curves apply to the fundamental power delivery capability of the generator
alone. They do not imply a given power output for any given electrode and cable configuration
when used with the generator. Each accessory will self-impose an upper set power limit for the
generator, the value of which will be equal to or below the maxium power delivery capability of
the generator.