Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The VAPR GENERATOR is intended for use in the electromagnetic environment specified below. The
customer or the user of the VAPR GENERATOR should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic Environment -
Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 Vrms
80 MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile RF communications
equipment should be used no closer to
any part of the VAPR GENERATOR,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended Separation Distance
d
= [1.17]√
P
d
= [1.17]√
P
80 MHz to 800 MHz
d
= [2.33]√
P
800 MHz to 2.5 GHz
Where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity
of equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in
which the VAPR GENERATOR is used exceeds the applicable RF compliance level above, the VAPR
GENERATOR should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the VAPR GENERATOR.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
EMC T
ablE
B-2