Recommended Separation Distances between Portable and Mobile RF Communications
Equipment and the VAPR GENERATOR
The VAPR GENERATOR is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the VAPR GENERATOR can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the VAPR GENERATOR as recommended below,
according to the maximum output power of the communications equipment.
Rated Maximum
Output Power of
Transmitter
W
Separation Distance According to
Frequency of Transmitter
m
150 kHz to 80 MHz
d
= [3.5/V1]√
P
80 MHz to 800 MHz
d
= [3.5/E1]√
P
800 MHz to 2.5 GHz
d
= [7/E1]√
P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
EMC T
ablE
B-3
MEDICAL EQUIPMENT
WITH RESPECT TO ELECTRICAL
SHOCK, FIRE AND MECHANICAL
HAZARDS ONLY IN ACCORDANCE
WITH UL2601-1 CAN/CSA C22.2 No
601.1. 21FA E176665
VAPR 3 Only