uidance and Manufacturer's Declaration - Electromagnetic Emissions
The VAPR GENERATOR is intended for use in the electromagnetic environment specified
below. The customer or the user of the VAPR GENERATOR should assure that it is used in
such an environment.
Emissions Test
RF Emissions
CISPR 11
Compliance
Group 1
Electromagnetic Environment - Guidance
The VAPR GENERATOR uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class A
Harmonic Emissions
IEC 61000-3-2
Class A
The VAPR GENERATOR is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Voltage Fluctuations /
Flicker Emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The VAPR GENERATOR is intended for use in the electromagnetic environment specified below. The
customer or the user of the VAPR GENERATOR should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level Electromagnetic Environment -
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV Contact
±8 kV Air
±6 kV Contact
±8 kV Air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical Fast
Transient / Burst
IEC 61000-4-4
±2 kV for Power
Supply Lines
±1 kV for Input /
Output Lines
±2 kV for Power
Supply Lines
±1 kV for Input /
Output Lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV Differential
Mode
±2 kV Common
Mode
±1 kV Differential
Mode
±2 kV Common
Mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage Dips, Short
Interruptions and
Voltage Variations
on Power Supply
Input Lines
IEC 61000-4-11
<5 %
UT
(>95 % Dip in
UT
)
for 0.5 Cycle
40 %
UT
(60 % Dip in
UT
)
for 5 Cycle
70 %
UT
(30 % Dip in
UT
)
for 25 Cycle
<5 %
UT
(>95 % Dip in
UT
)
for 5 sec
<5 %
UT
(>95 % Dip in
UT
)
for 0.5 Cycle
40 %
UT
(60 % Dip in
UT
)
for 5 Cycle
70 %
UT
(30 % Dip in
UT
)
for 25 Cycle
<5 %
UT
(>95 % Dip in
UT
)
for 5 sec
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the VAPR
GENERATOR requires continued
operation during power mains
interruptions, it is recommended that
the VAPR GENERATOR be powered
from an uninterruptible power supply or
a battery.
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Appendix B
EMC T
ablE
EMC T
ablE
B-1