8-2
If this does not resolve the problem, contact DePuy Mitek customer service.
A list of all Fault symbols and their descriptions is provided below:
There are no user defined alarms used in the VAPR system.
There are no distributed alarms used in the VAPR system.
There is no facility for disabling or silencing alarms used in the VAPR system.
All alarm conditions produced by the VAPR system are LOW PRIORITY technical alarms as
per IEC60601-1-8 table 1.
Audio alarms comply with the requirements of IEC 60601-2-2:2009.
Thermistor Electrode Alarm Descriptions
Short start: 500Hz for 75mS followed by 75mS gap repeated five times.
Short end: 500Hz for 130mS followed by 130mS gap repeated three times.
Tip Over temperature: 800Hz for 110mS followed by 500Hz for 110mS repeated while
condition occurs.
Non-Thermistor Electrode Alarm Description.
Short 65ms at 800Hz followed by 65ms at 500Hz while condition occurs and pedal pressed.
Fault
Symbol Ref Description
100
10
Software failure
100
11
ROM checksum failure
100
12
Non volatile memory failure
100
13
Memory failure
100
14
Power generation fault on start up
100
15
Power generation shutdown fault
200
10
PWM signal : shorted high
200
11
PWM signal : shorted low
200
12
MSYNC signal : shorted high
200
13
MSYNC signal : shorted low
200
14
Energy stuck high
200
15
IOUT stuck high
200
16
FOOT signal : shorted high
200
17
FOOT signal : shorted low
200
18
CUT/COAG signal : stuck high
200
19
CUT/COAG signal : stuck low
200
20
CLAMP signal error (DAC output)
200
21
BOOST signal error (DAC output)
200
22
PEAKSET signal error
200
23
PEAK signal error : stuck high
200
24
RF_DET signal error : stuck low
200
25
OVERDOSE signal error : stuck high
200
26
ENERGY signal error : stuck low
200
27
Line voltage relay error