CLEANING THE GENERATOR
WARNING
Electric shock hazard: Turn off the Generator and unplug the power cord from the AC source
prior to cleaning Generator.
The VAPR Generator cannot be sterilized. The Generator surfaces can be cleaned with a non-
abrasive cleaning agent. Do not allow fluids to enter the Generator connectors.
CLEANING THE FOOTSWITCH
The VAPR Footswitch cannot be sterilized. The Footswitch surfaces can be cleaned with
detergents and disinfectant cleaners according to standard hospital practices.
NOTE
The use of strong alkali detergents or cleaners must be avoided as these may damage the device.
CLEANING AND STERILIZING THE VAPR HANDPIECE
AND CABLE
1. Remove all gross matter (blood, mucous, tissue) by wiping each component with a
cloth or gauze pad and a mild cleaning solution or blood-dissolving detergent.
NOTE
The accessories are delicate surgical instruments. Do not immerse in reprocessing solutions. Do
not use abrasive cleaning agents. Product damage may otherwise result.
2. Rinse thoroughly in running water.
3. Allow the Handpiece to drain thoroughly.
4. Remove residual cleansing agents with a damp cloth.
5. Dry the accessory devices thoroughly before sterilizing.
STEAM STERILIZATION PROCEDURE
• Wrapped Pre-Vacuum Cycle: 134 to 136°C for 3 to 4 minutes.
• Drying times may vary with the type of wrapping material. Please refer to the sterilizer
manufacturer for recommended drying times.
• After any sterilization and cleaning process, check the Handpiece for any damage
(e.g. crush damage, cracking, or distortion). Discard if damage is observed.
WARNING
The DePuy Mitek Handpiece and Cable is intended for 20 reuse cycles only. Exceeding the
recommended number of uses may result in electrical or mechanical failure during use or
difficulty when assembling or disassembling the Electrode with the Handpiece.