Problem
Suggestions/Solutions
Output Shorted
(VAPR 3 ONLY)
When working in a joint cavity the electrode
tip may make contact with metallic
instrumentation close to the operative site.
As a safety measure, the generator will
immediately terminate activation and show a
warning message with a short alarm tone.
In the case of any short, remove the electrode
from the joint and inspect the accessories for
damage.
If the foot pedal remains depressed activation
will automatically restart following a short
period. For VAPR 3 variants with older V1.01
software and/or the use of TC electrodes,
the 'Output Shorted' fault may be cleared by
depressing either the footswitch or generator
mode buttons to restore the system to its
ready condition.
Check the accessories by activating the
electrode immersed in irrigating fluid contained
in a bowl, or similar, remote from patient
contact before proceeding with surgery. If the
fault recurs, first replace the electrode and
check as above. If the fault remains, replace
the hand piece and check as above.
Do not use any accessory that persistently
faults.
If the fault continues, contact DePuy Mitek
Customer Service, and discontinue use of
accessory.
Unable to access temperature control mode
The Temperature Control mode is only
enabled with purpose built TC electrodes with
an internal temperature sensing element.
Footswitch inoperative
Check the activation source has not been set
to HANDPIECE.
(VAPR II ONLY)
Generator Displays
“UNSUPPORTED TYPE” after an electrode
is inserted
Contact customer service to discuss
an upgrade to a new generator model.
(VAPR 3 ONLY)