background image

 

Spirodoc 

 

 

cod. 980156 

 Rev 1.3    

 

EN 7/67 

 

E

N

GLI

S

H

 

SPIRODOC

  is  continuously  checked  during  manufacturing  and  therefore  the  product 

complies  with  the  established  security  levels  and  quality  standards  laid  down  by  the 
Council Directive 93/42/EEC for MEDICAL DEVICES. 
After  removing  the  device  from  its  packaging,  check  to  see  that  there  is  no  visible 
damage. In case of damage do not use the device and return it to the manufacturer for 
repair. 

WARNING 

  

The safety and the correct performance of the device can only be assured if 
the user respects all of the relevant safety rules and regulations.  
The  manufacturer  will  not  be  held  responsible  for  damage  due  to  user

’s 

neglect to correctly to follow these instructions.  
The device must be used only and exclusively as a spirometer following the 
indications  given  by  the  manufacturer  with  particular  attention  to  the 
paragraph  on  INTENDED  USE,  and  utilizing  only  original  spare  parts  and 
accessories.  Use  of  non-original  parts  such  as  the  turbine  flow  sensor  and 
oximetry  sensor  or  other  accessories  may  cause  errors  in  measurement 
and/or compromise the correct functioning of the device, and is therefore not 
permitted. 
The device should not be used beyond the declared life span. The life span is 
strictly  related  to  the  life  of  the  internal  lithium  battery  pack.  In  normal 
conditions  the  lifespan  of  the  battery  pack  is  estimated  to  be  around  10 
years. The device constantly monitors the state of charge of this battery and 
a message informs the user when the battery is discharged.  
In the event of any incident or accident of any kind resulting from the use of 
the device, the user is required to inform the manufacturer without delay, this 
procedure is laid down in Article.9 of the European Regulations No. 46/1997, 
which implemented the EC Directive No. 93/42. 

 

2.2.1 

Danger of cross-contamination 

Two  different  types  of  turbine  sensors  can  be  used  with  the  device,  one  is  reusable 
and  the  other  is  single-patient  disposable.  A  disposable  mouthpiece  is  required  in 
order to connect a  patient to the spirometer. In order to avoid exposing the  patient to 
the  critical  danger  of  cross-contamination,  the  reusable  flow  sensor  must  always  be 
cleaned  before  each  spirometry  test,  and  a  new  disposable  mouthpiece  must  always 
be  used  for  each  patient.  The  use  of  an  anti-bacterial  filter  is  at  the  discretion  of  the 
doctor. If a single-patient disposable turbine is used, then a new one must be used for 
each patient. 
 

2.2.2 

Turbine 

 

Disposable turbine

 

 

WARNING 

If  you  are  going  to  perform  a  spirometry  test  with  a 
disposable  turbine  it  is  important  to  use  a  new  turbine 
for  each  new  patient.  The  features,  accuracy  and  
hygiene  of  the  disposable  turbine  can  only  be 
guaranteed  if  it  has  been  conserved  beforehand  in  its 
original sealed packaging. 

Summary of Contents for Spirodoc

Page 1: ...tery pack warning 11 2 5 Labels and symbols 14 2 5 1 Identification label 14 2 5 2 CE mark for medical devices 14 2 5 3 Electrical safety symbol 14 2 5 4 Warning symbol for the RS232 serial port 14 2...

Page 2: ...gle Patient Sensor 49 4 datA TRANSMISSION 50 4 1 Data Transmission via Bluetooth to a cell phone 51 4 1 1 Preliminary Operations 51 4 1 2 Bluetooth data transmission 51 4 2 Data transmission via Bluet...

Page 3: ...001 3 disposable turbines 910001 1 oximetry sensor 919010 Accessori COD Accessori COD Battery charger with micro USB connector 920680 Battery charger with micro USB connector 920680 Disposable adult o...

Page 4: ...d or comments regarding this product may be sent via email to mir spirometry com MIR accepts no responsibility for any loss or damage caused by the user of the device due to instructions contained in...

Page 5: ...C has been designed for use in the doctor s office in a hospital or directly by the patient to continuously monitor her his physical conditions during routine daily activities All information necessar...

Page 6: ...before a spirometry test is made The user is responsible to assess both the mental and the physical condition of the patient in order to perform a proper test furthermore in the evaluation of test res...

Page 7: ...thium battery pack In normal conditions the lifespan of the battery pack is estimated to be around 10 years The device constantly monitors the state of charge of this battery and a message informs the...

Page 8: ...ce of any impurities such as hair sputum threads etc within the body of the turbine sensor may seriously compromise the accuracy of the measurements 2 2 3 Mouthpiece Any disposable mouthpieces include...

Page 9: ...sensors supplied with or specifically intended for use with SPIRODOC Use of oximetry sensors not intended for use with the SPIRODOC may cause inaccurate readings Oximetry measurements may be inaccurat...

Page 10: ...h as a TV radio portable phone etc and other electronic units are operated at the same time in the same room The device may give inaccurate readings if operated in the presence of strong electromagnet...

Page 11: ...lied with the device which delivers 3 7 volts For proper use of the battery pack please read carefully the warning below WARNING Do not use the battery pack for any purposes different from those speci...

Page 12: ...fire Do not use or store the battery near a fire or if the temperature inside a vehicle may reach 60 C or become higher Do not immerge the battery pack in water or salt water and do not leave it wet S...

Page 13: ...ore using the battery pack carefully read the manual paying attention to all the recommendations for proper handling For information concerning the installation and removal of the battery pack careful...

Page 14: ...duct is certified to conform to the Class II requirements of the 93 42 EEC medical device directive 2 5 3 Electrical safety symbol In accordance with the IEC 60601 1 Standard this product and its comp...

Page 15: ...following conditions 1 this device must not cause harmful interference 2 this device must accept any interference received including interference that may cause undesired operation Any modifications n...

Page 16: ...rement device intended for use by a respiratory specialist or by a suitably trained general practitioner The spirometer calculates up to 30 functional respiratory parameters providing the pharmacodyna...

Page 17: ...ivo utilizzando la connessione Bluetooth al PC del medico Pertanto il medico pu valutare a distanza i parametri legati alla patologia del paziente SPIRODOC can also be connected to a PC or to another...

Page 18: ...test L FEV1 FVC FEV1 FVC x 100 FEV1 VC FEV1 best between EVC and IVC x 100 PEF Peak expiratory flow L s FEF2575 Average flow between 25 and 75 of the FVC L s FEF25 Forced Expiratory Flow at 25 of FVC...

Page 19: ...try SYMBOL DESCRIPTION Units SpO2 Baseline SpO2 Average in first three minutes SpO2 Min SpO2 Minimum during period of analysis SpO2 Max SpO2 Maximum during period of analysis SpO2 Mean SpO2 Average du...

Page 20: ...above 5 minutes hh mm ss NOD 90 Time passed with SpO2 90 for continued periods above 5 minutes with minimum value 86 Nadir hh mm ss DELTA Parameters measured for six minute walk test analysis SYMBOL...

Page 21: ...rea under SpO2 curve base relative to distance covered Dyspnea Borg CHG Variation in grade of dyspnea during walking Fatigue Borg CHG Variations in level of fatigue during walking DELTA There follows...

Page 22: ...frared absorption Range of measurement SpO2 0 100 with 1 increments SpO2 Resolution 1 SpO2 accuracy 2 between 70 100 SpO2 Average number of heart beats for the SpO2 calculation 8 beats Range of measur...

Page 23: ...l protection BF Grade of protection against water ingress IPX1 device protected against water drops Safety level in the presence of inflammable anaesthetic gas oxygen or nitrogen Device not suitable C...

Page 24: ...access the default settings service menu To access patient data from the main display To perform a new test of a patient recalled from the patient records To insert new patient data To modify patient...

Page 25: ...on after approximately 10 seconds of disuse the display enters energy saving mode thereby lowering the display contrast level automatically If the device remains in disuse for approximately 5 minutes...

Page 26: ...Date Time Date Format Unit Format Turbine selection Standards setting LCD setting BLUETOOTH setting Firmware information Oximetry alarms SpO2 sampling rate Spirodoc mode Parameter settings Menu type S...

Page 27: ...rom left to right 1 2 2 3 3 3 If an error occured during insertion of the password the following message will appear Password Error Press OK and try again If you fail to insert the password again the...

Page 28: ...HANES III The selection will be saved automatically and the device will return to the service menu LCD setting This menu allows for Setting of display luminance and contrast Two scales ranging from 0...

Page 29: ...onents Spirodoc Bluetooth After approximately 10 seconds the device automatically displays the Service Menu otherwise press Oximetry Alarms Access to this function allows the setup of reference values...

Page 30: ...plified mode the user can only view the main parameters as per actual standards The personalized mode allows to select the parameters of interest The complete mode allows to view all the parameters th...

Page 31: ...rd To access the service menu select the item Inser password and enter the following password 1 2 2 3 3 3 In this mode the service menu displays the following items Spirodoc mode Personal best Turbine...

Page 32: ...e chosen by the patient Domanda Possibili risposte Tiredness on waking NO MED MAX Daytime drowsiness NO MED MAX Breathless on waking NO MED MAX Troubled sleep NO MED MAX Wheezing NO MED MAX Cough NO M...

Page 33: ...these are displayed next to the New FVC and New FIVC parameters The message ENTER OK TO CONFIRM will appear By touching the ESC icon the device will automatically return to the service menu without a...

Page 34: ...ibrated correctly then the FVC syringe value will be 3 00 FVC x 1 026 BTPS 3 08 L FVC at BTPS If the ambient temperature is 20 C the FIVC syringe value will be 3 00 FIVC x 1 102 BTPS 3 31 L FIVC at BT...

Page 35: ...n Oriental 100 Afro American Hong Kong Chinese 100 Other Giapanese 89 polinesian 90 North Indian 90 South Indian 87 Pakistani 90 African descendant 87 For ATS ERS standards the correction is applied t...

Page 36: ...ouch the OK icon The data returned by the device are all the test sessions performed during that specific day Complete database shows data starting from the most recent session The end of the database...

Page 37: ...sage EMPTY LIST Device search Touch OK to confirm By touching the OK icon the device will search for any nearby devices if any devices are found these can be memorized in the specific Bluetooth list 3...

Page 38: ...er to perform proper spirometry testing the following instructions are to be followed carefully Insert the turbine in the appropriate housing until it reaches the mechanic stop and successively rotate...

Page 39: ...nspiratory parameters FIVC FIV1 FIV1 PIF are not of interest The optional initial inspiration phase can also be performed before inserting the mouthpiece in the mouth After inspiring slowly and deeply...

Page 40: ...iratory Slow Vital Capacity VC After the acoustic signal inspire slowly as much as air as possible and expire slowly as much air as possible Inspiratory Slow Vital Capacity IVC After the acoustic sign...

Page 41: ...sion a new patient is inserted or another is recalled from the archive the device will automatically exit the current POST session 3 12 Viewing the spirometric results Following a FVC test the spirome...

Page 42: ...more than one but the FEV1 values not matching to within 200 mL with no interpretation F No acceptable manoeuvres with no interpretation POST test A two acceptable 1 FEV1 values matching within 100 m...

Page 43: ...RNING If SPIRODOC has been purchased without the oximetry option only spirometry testing can be performed If the oximetry option is purchased afterwards please contact a service center or the manufact...

Page 44: ...probe Ensure that the bottom part of the finger completely covers the detector If the finger cannot be placed properly inside the sensor try another finger Position the sensor so that the cable is und...

Page 45: ...itself and the device starts recording data For any type of oximetry test if the sensor is not properly connected the following message will be displayed on screen after a few seconds WARNING Sensor n...

Page 46: ...formation concerning the proper setup of this function please refer to paragraph 3 6 WARNING A test is saved with the name of the last patient displayed If a test refers to a previously saved patient...

Page 47: ...the timer is reset to zero so that the user can immediately see of the duration of each single phase The data on the display is the same as shown before This stage will continue for a minimum of 2 min...

Page 48: ...calculated during the initial stage of the test 3 13 2 Sleep Oximetry To access the oximetry area from the main screen touch the icon next select the icon to select the test This test records overnigh...

Page 49: ...refer to paragraph 4 2 for a data printout an example can be found in the attachments inside this user manual 3 13 4 Instructions for Adult Single Patient Sensor WARNING The oximetry sensor used in t...

Page 50: ...ector Wrap the adhesive around the finger or toe to secure the sensor Route the cable along the palm or the bottom of the foot and secure with adhesive tape if necessary Connect the sensor to the devi...

Page 51: ...n on the SPIRODOC Touch the icon on the main screen The type of transmission selected is shown if this is correct then confirm with OK to activate the connection to the default device that was initial...

Page 52: ...est of interest was performed Upon entering the test session select the test and touch the icon If no device was setup to print via Bluetooth a message will appear on screen which allows the device to...

Page 53: ...matic driver installation for Windows 98 2000 ME or request some information for Windows XP Vista and Seven To avoid making any errors at this stage please read the Advanced section of the winspiroPRO...

Page 54: ...advantage of requiring no periodic calibration In order to maintain the default characteristics of the reusable turbine a simple cleaning procedure is required before use Cleaning of the disposable tu...

Page 55: ...s housing making sure to turn it clockwise as shown by the symbol of the lock printed on SPIRODOC The turbine is inserted properly by pushing it all the way in and subsequently rotating it clockwise u...

Page 56: ...on SPIRODOC and the following icon will appear on the main screen showing the charge level of the battery pack The maximum charge level is displayed with all 6 bars inside the battery If only one bar...

Page 57: ...evice by using the winspiroPRO software For further information please refer to the winspiroPRO user manual available inside the software itself At the end of a spirometry test the data measurements a...

Page 58: ...No 0476 Certificate No MED 9826 MIR srl Medical International Research declares that the Device subject of this declaration together with its standard accessories conforms to the requirements of the C...

Page 59: ...oes not conform to the current safety norms in the country of installation If the product is utilised differently from the use described in the User s Manual If any alteration adjustment modification...

Page 60: ...60 67 ENGLISH MIR Medical International Research reserves the right to modify the device if required and a description of any modification made will be sent along with the returned goods ALLEGATO 1 O...

Page 61: ...Spirodoc cod 980156 Rev 1 3 EN 61 67 ENGLISH...

Page 62: ...Spirodoc cod 980156 Rev 1 3 EN 62 67 ENGLISH...

Page 63: ...Spirodoc cod 980156 Rev 1 3 EN 63 67 ENGLISH ALLEGATO 2 SPIROMETRY TEST REPORT EXAMPLE...

Page 64: ...ser of the Spirobank II should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The SPIRODOC uses RF energy on...

Page 65: ...mmon mode Not Applicable Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 6100...

Page 66: ...res objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broa...

Page 67: ...smitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 3 5 P 3 80 MHz to 800 MHz d 3 5 P 3 800 MHz to 2 5 GHz d 7 P 3 0 01 0 12 0 24 0 24 0 1 0 37 0 37 0 74 1 1 17 1...

Reviews: