Spirodoc
cod. 980156
Rev 1.3
EN 19/67
E
N
GLI
S
H
TV/t
I
Average flow of inspiration, at rest
L/min
t
I
/Ttot
t
E
/(t
I
+t
E
)
\
MVV
Maximum voluntary ventilation
L/min
ELA
Estimated lung age
year
*= best values
Flow/volume measurement system
Bi-directional digital turbine
Temperature sensor
semiconductor (0-45°C)
Measurement principle
Infrared interruption
Volume range
10 L
Flow range
16 L/s
Volume accuracy
3% or 50 mL
Flow accuracy
5% or 200 mL/s
Dynamic resistance at 12 L/s
<0.5 cmH
2
O/L/s
2.6.2
Oximeter features
Definitions:
Desaturation Event
Desaturation events SpO2 fall
4% in a limited period of 8-40 sec
and successive rise
2% within a total period of 150 sec.
Total Pulse rate
Variation
Pulse rate rise
10 BPM in limited period of 8-40 sec and
successive fall
8 BPM during a total period of 150 sec.
Parameters measured during sleep oximetry:
SYMBOL
DESCRIPTION
Units
SpO2 Baseline
SpO2 Average in first three minutes
%
SpO2 Min
SpO2 Minimum during period of analysis
%
SpO2 Max
SpO2 Maximum during period of analysis
%
SpO2 Mean
SpO2 Average during period of analysis
%
BPM Baseline
Average pulse frequency in the first 3 minutes
BPM
BPM Min
Minimum pulse frequency during the period of
analysis
BPM
BPM Max
Maximum pulse frequency during the period of
analysis
BPM
BPM Mean
Average pulse frequency during the period of
analysis
BPM
Recording time
Total time measure of SpO2
hh:mm:ss
T < 90%
Time passed with SpO2 < 90 %
% hh:mm:ss
T < 89%
Time passed with SpO2 < 89 %
% hh:mm:ss
T < 88%
Time passed with SpO2 < 88 %
% hh:mm:ss
T < 87%
Time passed with SpO2 < 87 %
% hh:mm:ss
N° Events SpO2
<89%
Fall of SpO2 below 89% for at least 20 seconds
\
Index [12s]
Index of SpO2 fluctuation calculated in intervals
of 12 seconds
\
T< 40 BPM
Time passed with pulse frequency < 40 BPM
% hh:mm:ss
T> 120 BPM
Time passed with pulse frequency > 120 BPM
% hh:mm:ss
N° Events < 40
BPM
Bradycardia events during the entire period of
analysis
\