Spirodoc
cod. 980156
Rev 1.3
EN 57/67
E
N
GLI
S
H
- The battery pack could be completely discharged; recharge the device by connecting
it to the battery charger.
- The device internal software is lost. Connect the device to the PC via USB and
proceed to download the internal software. For further information please refer to the
winspiroPRO user manual available inside the software itself.
While in use the device turns off and on again.
- An internal error has occured, please check on the website www.spirometry.com for
new internal software upgrades. Should there be a newer internal software version
please upgrade the device by using the winspiroPRO software. For further information
please refer to the winspiroPRO user manual available inside the software itself.
At the end of a spirometry test the data measurements are not reliable.
- Clean the turbine as shown in paragraph 5.1; if necessary replace the turbine with a
new one.
Memory loss due to an unexpected event
- The database has been lost. Contact a local technical service center.
WARNING
Before contacting a technical assistance center, please try downloading the
database from the device to the PC using the winspiroPRO software. This
procedure is necessary as a backup in case all the data is accidentally lost
during device repair. Furthermore the database could be of confidential nature
and as such not accessible by authorized personnel and also subject to privacy
laws.