Trio™ Service Manual
0070-10-0591-01
5 - 3
Appendix
System Alarm Prompts
*XX RANGE EXCEEDED
The measured value of XX
parameter has exceeded the
system measurement range
If message does not clear
contact Customer Service.
REAL CLOCK NEEDSET
If the monitor displays 2000-1-
1, this system prompt reminds
the user that the current system
time is inaccurate
Reset the system time. After
setting the time, the user
should cycle power. This will
avoid any time storage
errors.
REAL CLOCK NOT EXIST
The system has no cell battery
or the battery has run down
Install or replace the cell
battery
SYSTEM WD FAILURE
The system has a fatal error or
failure
Restart the system. If message
does not clear contact
Customer Service.
SYSTEM SOFTWARE ERR
SYSTEM CMOS FULL
SYSTEM CMOS ERR
SYSTEM EPGA FAILURE
SYSTEM FAILURE2
SYSTEM FAILURE3
SYSTEM FAILURE4
SYSTEM FAILURE5
SYSTEM FAILURE6
SYSTEM FAILURE7
SYSTEM FAILURE8
SYSTEM FAILURE9
SYSTEM FAILURE10
SYSTEM FAILURE11
SYSTEM FAILURE12
KEYBOARD NOT AVAILABLE
The keypad cannot be used
Check the keys for abnormal
depression. Contact the
manufacturer for repair.
KEYBOARD COMM ERR
The keypad has a failure
Contact the manufacturer for
repair
KEYBOARD ERROR
KEYBOARD FAILURE
KEYBOARD ERR1
KEYBOARD ERR2
5V TOO HIGH
The power supply has a failure
Restart the system. If message
does not clear contact
Customer Service.
5V TOO LOW
POWER ERR3
POWER ERR4
12V TOO HIGH
12V TOO LOW
POWER ERR7
POWER ERR8
3.3V TOO HIGH
3.3V TOO LOW
PROBLEM/MESSAGE
REASON
SOLUTION
*
XX represents all the parameter modules in the system such as ECG, NIBP, SpO
2
, IBP module, etc.