Trio™ Service Manual
0070-10-0591-01
iii
Foreword
Introduction
Foreword
This service manual gives a detailed description of the
Trio
Portable Patient Monitor,
including, circuit descriptions, test procedures and a spare part listing. This manual is
intended as a guide for technically qualified personnel during repair, testing or calibration
procedures.
Warnings, Precautions And Notes
Please read and adhere to all warnings, precautions and notes listed here and in the
appropriate areas throughout this manual.
A
WARNING
is provided to alert the user to potential serious outcomes (death, injury, or
serious adverse events) to the patient or the user.
A
CAUTION
is provided to alert the user to use special care necessary for the safe and
effective use of the device. They may include actions to be taken to avoid effects on patients
or users that may not be potentially life threatening or result in serious injury, but about which
the user should be aware. Cautions are also provided to alert the user to adverse effects on
this device of use or misuse and the care necessary to avoid such effects.
A
NOTE
is provided when additional general information is applicable.
Warning
WARNING: The NIBP pneumatic test (specified in the EN 1060-1
standard) is used to determine if there are air leaks in the
NIBP airway. If the system displays the prompt that the
NIBP airway has air leaks, please contact the manufacturer
for repair.
CAUTION:
To ensure continued use of the Factory Defaults when the
unit is powered off and on, save the Factory Defaults as the
User Default Configuration after reassembly.