Trio™ Service Manual
0070-10-0591-01
2 - 1
2.0
Calibration/Maintenance
2.1
Calibration Introduction
The following procedures are provided to verify the proper operation of the
Trio
Monitor. A
menu driven interface is used to execute all verification tests.
2.2
Warnings and Guidelines
In the event that the instrument cover is removed, observe these following warnings and
general guidelines:
1.
Do not short component leads together.
2.
Perform all steps in the exact order given.
3.
Use extreme care when reaching inside the opened instrument. Do not contact exposed
metal parts which may become ‘‘live’’.
4.
Read through each step in the procedure, so it is understood prior to performing the
step.
2.3
Test Equipment and Special Tools Required
• Digital or Mercury Manometer w/bulb and valve 0-300 mmHg
• Test Chamber/Dummy Cuff (P/N 0138-00-0001-03)
• DVM
• Patient Simulator
• Safety Analyzer (Dempsy Model 431 or Equivalent)
• Metric Ruler
• PC or Laptop w/Windows 98 or above, CD-ROM drive, and Ethernet card installed