System Alarm Prompts
Appendix
5 - 4
0070-10-0591-01
Trio™ Service Manual
CELL BAT TOO HIGH
Cell battery is defective.
Incorrect cell battery installed
Replace the battery. If
message does not clear
contact Customer Service.
CELL BAT TOO LOW
The cell battery has become
depleted. The cell battery is not
installed or the connection is
loose.
RECORDER SELF TEST ERR
During the recorder self test,
there is a Recorder Module
communication failure
Execute
Clear Record Task
function in the recorder setup
menu to reset communication
between the host and the
recorder. If message does not
clear contact Customer
Service.
RECORDER VLT HIGH
The recorder module has
voltage failure
Contact the manufacturer for
repair
RECORDER VLT LOW
RECORDER HEAD HOT
The continuous recording time
may be too long
After the recorder cools
down, use again. If the
problem still exists, contact
the manufacturer for repair.
REC HEAD IN WRONG
POSITION
The recorder roller lever is not
pressed down
Press down the recorder roller
lever
RECORDER OUT OF PAPER
No paper is in the recorder
Install paper into the recorder
RECORDER PAPER JAM
The paper in the recorder is
jammed
Re-install the recorder paper
and try again
RECORDER COMM ERR
The communication of the
recorder is abnormal
Execute
Clear Record
Task
from the recorder setup menu
to reset communication
between the host and the
recorder. If message does not
clear contact Customer
Service.
RECORDER PAPER W.P.
The recorder paper is not
installed in the correction
position
Install the recorder paper in
the correct position
REC NOT AVAILABLE
Cannot communicate with the
recorder
Execute
Clear Record Task
from the recorder setup menu,
to reset communication
between the host and the
recorder. If message does not
clear contact Customer
Service.
NIBP INIT ERR
NIBP initialization error
Restart the system. If message
does not clear contact
Customer Service.
NIBP SELFTEST ERR
NIBP ILLEGALLY RESET
During NIBP measurement,
illegal reset occurs
Check the airway of NIBP to
see if there are occlusions.
Then measure again. If the
problem persists, contact the
manufacturer for repair.
NIBP COMM ERR
No NIBP communication
Restart the system. If message
does not clear contact
Customer Service.
PROBLEM/MESSAGE
REASON
SOLUTION
*
XX represents all the parameter modules in the system such as ECG, NIBP, SpO
2
, IBP module, etc.