4.7.1 Basic
Figure 4-19 Basic setup of the system information
The following table explains the parameters on the
Basic
screen.
Parameter Description
FPGA
Software
The version information of the system.
Hospital
Name of the hospital.
Start button
sound
Start the function of making sound when pressing a button.
Pump
speed
3 pump speeds are available: 1, 2, 3. The default which is
recommended to the user is 1.When there are some impurities in
the flow cell or tubing that can not be cleared, select pump speed 2
or 3 to clear them. After washing, it is recommended that pump
speed 1 should be restored.
Date
Select the date in the date table. Use and
to select month and
year.
Time
Enter time in the text box.
Standby
interval
Set the time interval for the system to enter Standby status.
Colorimetri
c mode
Use cuvette or cell.