2-2
Ambient temperature: 15
℃
-30
℃
Relative humidity: 35%-85%, without condensation
Above-sea-level height: -400m
~
3,000 m
(
70kPa
~
1,06kPa
)
Power supply: 100V-240V~ (
±
10% fluctuation), 50/60Hz (
±
3Hz fluctuation).
The site should be away from the electromagnetic noise and the bearing
platform should be free of vibration.
The site should not be in direct sun
The system should be properly grounded.
WARNING
Please make sure that the power is properly grounded. Improper
grounding may lead to electric shock and/or equipment damage. Be
sure to connect the system to a power socket that meets the
above-mentioned requirements and has a proper fuse installed.
2.2.3 Space
and
Accessibility
Requirements
Figure 2-1 Space and accessibility requirements
The installation space must be larger than the dimensions (420mm × 350mm ×
158mm, L×W×H) of the instrument. Please refer to Figure 2-1.
The installation platform should be level with gradient less than 1:200.
2.2.4 Water
Requirements
1)
The water must meet requirements of the CAPII water.
2)
The water temperature should be within 15-5
0℃