4-8
4.2.1.4 QC Information
The
QC Information
screen is where you can set the test QC rule and control
information.
Figure 4-6 QC information
The following table explains the parameters on the
QC Information
screen.
Parameter Description
Rule
Set QC rule for the selected test. One or more rules can be
selected, or none is selected.
Control
You can set 3 controls of different concentrations and one QC rule
for each test.
For any level of control, once any one of the parameters (control
lot, concentration or SD) is entered, you should enter the rest of the
parameters to complete the setting. Otherwise, the control
information can not be saved, and error message will appear.
The following table introduces the buttons on the page.
Buttons
Function
Save
Save the test QC information.
Cancel
After selecting a test and setting or modifying parameters for it, click
this button to cancel the setting or modification.
Return
Click this button to return to the main screen.
Summary of Contents for BA-88A
Page 1: ...BA 88A Semi auto Chemistry Analyzer Operation Manual...
Page 2: ......
Page 6: ......
Page 18: ......
Page 21: ...III Appendix C External Printer C 1 Appendix D Supplies D 1...
Page 22: ......
Page 28: ......
Page 78: ......
Page 96: ......
Page 102: ......
Page 104: ......
Page 106: ......
Page 107: ...P N BA89 20 87038 1 0...