B-2
Light signal too
weak
ADL
Lamp is not turned on, or light
source degenerated, or light
path is obstructed, or lamp falls
off, or gain error occurs
Restart the system or refer to the
5.3.2. If the alarm continues
contact Mindray customer service
department or your local
distributor.
Dark current is
too high
DRH AD collection error or circuit
noise too strong.
Restart the system. If the alarm
continues contact Mindray
customer service department or
your local distributor.
Time to for the
system to
stabilize after
startup out of
range.
001
Temp. data collection error;
refrigeration plate error; main
board error; refrigeration plate
invalid.
Restart the system. If the alarm
continues contact Mindray
customer service department or
your local distributor.
Cuvette
temperature too
high
002 Temp. data collection error;
refrigeration plate error; main
board error; refrigeration plate
invalid.
Restart the system. If the alarm
continues contact Mindray
customer service department or
your local distributor.
Cuvette
temperature too
low
003 Temp. data collection error;
refrigeration plate error; main
board error; refrigeration plate
invalid.
Restart the system. If the alarm
continues contact Mindray
customer service department or
your local distributor.
Cuvette
temperature
fluctuation out
of range
004 Temp. data collection error;
refrigeration plate error; main
board error; refrigeration plate
invalid.
Restart the system. If the alarm
continues contact Mindray
customer service department or
your local distributor.
Fan blocked
during running
005
Fan blocked or damaged
Restart the system. If the alarm
continues contact Mindray
customer service department or
your local distributor.
Temperature
switching
exceeds time
limit
006 Temp. data collection error;
refrigeration plate error; main
board error; refrigeration plate
invalid.
Restart the system. If the alarm
continues contact Mindray
customer service department or
your local distributor.
Sensor
abnormal
007
Main board error or temp.
sensor abnormal or temp. out
of measurable range.
Restart the system. If the alarm
continues contact Mindray
customer service department or
your local distributor.
Filter wheel can
not find home
position
008
Photocoupler damaged or dust
on photocoupler or motor
damaged.
Restart the system. If the alarm
continues contact Mindray
customer service department or
your local distributor.
Light signal too
weak
009
Lamp is not turned on, or light
source degenerated, or light
path is obstructed, or lamp falls
off, or gain error occurs
Restart the system or refer to the
5.3.2. If the alarm continues
contact Mindray customer service
department or your local
distributor.
Photoelectric
collection is
abnormal
010
Simulation board error
Restart the system. If the alarm
continues contact Mindray
customer service department or
your local distributor.
Summary of Contents for BA-88A
Page 1: ...BA 88A Semi auto Chemistry Analyzer Operation Manual...
Page 2: ......
Page 6: ......
Page 18: ......
Page 21: ...III Appendix C External Printer C 1 Appendix D Supplies D 1...
Page 22: ......
Page 28: ......
Page 78: ......
Page 96: ......
Page 102: ......
Page 104: ......
Page 106: ......
Page 107: ...P N BA89 20 87038 1 0...