4-30
Parameter Description
Sample ID Sample ID of the patient. If you enter the screen by selecting a test
result, the sample ID can not be edited. The sample ID should not be
void. Otherwise, the patient information can not be saved.
The entered sample ID can not be the same with the existing sample
ID of the current day.
Type
The options are serum, plasma and urea, etc. You can add or delete
the types in
Setup
-
Dictionary
.
Age
Age of the patients. The units are Year, Month, Day.
Gender
Gender of the patient, including: Male, Female.
MRN
Medical record No. of the patient.
Sender
Name of the doctor who sent the sample to the clinical laboratory.
You can add or delete the names in
Setup
-
Dictionary
.
Tester
Name of the doctor who tested the sample. You can add or delete the
names in
Setup
-
Dictionary
.
Test Date
The date when sample is tested.
Send Date
The date when sample is sent to be tested.
Reviewed
by
Name of the doctor who reviewed the test report. You can add or
delete the names in
Setup
-
Dictionary
.
Remark
More information about the patient.
The following introduces the buttons on the screen.
Buttons
Function
Click this button to view or edit the information of last sample.
Click this button to view or edit the information of the next sample.
Save
Save the information on setting.
Cancel
Cancel all the editing and return to the status before editing.
Print the report of the selected sample.
Return
Click this button to go back to
Sample
screen.
4.7 Setup
You can set hospital name, system date and time, data dictionary, touchscreen
calibration and delete history data in
Setup
.
Summary of Contents for BA-88A
Page 1: ...BA 88A Semi auto Chemistry Analyzer Operation Manual...
Page 2: ......
Page 6: ......
Page 18: ......
Page 21: ...III Appendix C External Printer C 1 Appendix D Supplies D 1...
Page 22: ......
Page 28: ......
Page 78: ......
Page 96: ......
Page 102: ......
Page 104: ......
Page 106: ......
Page 107: ...P N BA89 20 87038 1 0...