I - 4
Operator’s Manual
I EMC Guidance and Manufacturer’s Declaration
Voltage dips, Short
interruptions and voltage
variation on power supply
input voltage
IEC 61000-4-11
0 % U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
0 % U
T
; 1 cycle
70% U
T
for 25/30 cycle
at 0°
0 % U
T
; 250/300 cycle
0 % U
T
; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
0 % U
T
; 1 cycle
70% U
T
for 25/30 cycle
at 0°
0 % U
T
; 250/300 cycle
Mains power quality
should be that of a typical
commercial or hospital
environment. If you
require continued
operation during power
mains interruptions, it is
recommended that our
product be powered from
an uninterruptible power
supply or a battery.
Power frequency (50/60
HZ) magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE:
U
T
is the A.C. mains voltage prior to application of the test level.
Table I-3
GUIDANCE AND MINDRAY DECLARATION-ELECTROMAGNETIC IMMUNITY
The system is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
IMMUNITY TEST
IEC 60601
TEST LEVEL
COMPLIANCE LEVEL ELECTROMAGNETIC
ENVIRONMENT-
GUIDANCE
Summary of Contents for Ana
Page 2: ......
Page 50: ...This page intentionally left blank...
Page 60: ...This page intentionally left blank...
Page 110: ...This page intentionally left blank...
Page 116: ...This page intentionally left blank...
Page 166: ...This page intentionally left blank...
Page 176: ...This page intentionally left blank...
Page 194: ...This page intentionally left blank...
Page 220: ...This page intentionally left blank...
Page 288: ...This page intentionally left blank...
Page 304: ...This page intentionally left blank...
Page 308: ...This page intentionally left blank...
Page 316: ...This page intentionally left blank...
Page 337: ......
Page 338: ...P N 046 018835 00 2 0...