16 - 56
Operator’s Manual
16 Probes and Biopsy
c.
Hold the needle for 1 or 2 seconds and then withdraw it quickly from the magnetizer.
16.4.4 Procedure
NOTE:
•
The Pan Zoom function is not supported on eSpacial Navi mode.
•
Do not freeze the ultrasound image during the procedure. Otherwise, biopsy guidance
information will disappear.
•
When the insertion angle of a needle relative to the skin surface exceeds 60°,
is displayed
on screen to notify users to adjust the angle. After the angle is smaller than 60°,
disappears.
Perform the following procedure:
1.
Hold the probe stably and put the needle close to the probe for connection. After the needle and
probe are successfully connected, the system automatically calculate the maximum depth at
which the needle can be detected in real time, and display the depth value in the touch screen.
When the actual depth is greater than the maximum value, the needle guidance information
disappears temporally.
2.
Adjust the needle position and angle to define the position for inserting the needle.
Pay attention to the field strength during the needle guidance process.
16.4.5 Magnetizer Cleaning and Disinfection
After use, clean and disinfect the magnetizer as required. Failure to do so may result in the
magnetizer becoming sources of infection.
CAUTION
•
When performing cleaning and disinfection of the magnetizer to prevent
infection, wear sterile gloves.
•
After disinfection, rinse the magnetizer thoroughly with sterile water to
remove all chemical residues.
•
No cleaning and disinfecting may result in the magnetizer becoming a
source of infection.
•
Please follow the disinfectant manufacturer's manual for performing
cleaning and disinfection, including preparing sterile water and cleaning
and disinfection time.
NOTE:
•
Please refer to the instructions provided by the manufacturer of the disinfectant for details.
•
The magnetizer can be disinfected using the compatible disinfectants at least 500 times.
Summary of Contents for Ana
Page 2: ......
Page 50: ...This page intentionally left blank...
Page 60: ...This page intentionally left blank...
Page 110: ...This page intentionally left blank...
Page 116: ...This page intentionally left blank...
Page 166: ...This page intentionally left blank...
Page 176: ...This page intentionally left blank...
Page 194: ...This page intentionally left blank...
Page 220: ...This page intentionally left blank...
Page 288: ...This page intentionally left blank...
Page 304: ...This page intentionally left blank...
Page 308: ...This page intentionally left blank...
Page 316: ...This page intentionally left blank...
Page 337: ......
Page 338: ...P N 046 018835 00 2 0...