16 Probes and Biopsy
Operator’s Manual
16 - 25
2.
Adjust the needle angle to the proper shift as required:
a.
Loosen the angle pinch nut.
b.
Adjust the angle block to the desired level.
c.
Tighten the angle pinch nut.
3.
Install the guiding block
a.
Select a suitable guiding block and push it into the groove above the angle block, then
clamp it tightly
b.
Screw the block's nut to secure the block.
c.
Insert a biopsy needle with the same specification as that of the guiding block into the
guiding block hole.
4.
Release the needle from the bracket:
a.
Screw the nut of the guiding block and remove the guiding block slightly along the
direction of the needle’s tail.
b.
Separate the residual part of the needle-guide bracket and the probe from the needle.