E103 (run)
load cell
check wiring
E104 (run)
memory checksum
contact Mass Dynamics or
their distributor
OF
no speed signal
check speed circuit or
run conveyor
» Error messages are cleared when the condition is remedied.
» Run mode errors are suppressed during program mode,
zero or span calibration.
» Zero and span errors are cleared when zero or span is initiated. Errors
that happen during the calibration require re-starting the procedure.
» Program errors are cleared when any key is pressed.
» The messages E101 through E104 can be turned off (P949).
» The load cell errors are detected by certain conditions where it is apparent
that the load cell is not functioning or incorrectly wired. It is not a conclusive
test, since even with incorrect wiring, the resulting input from the load cell
may be within a valid range.
APPENDICES
PL-531
81