eGROSS pro-x - eGROSS
MM137-004 - Operator Manual
32
2.3.3. Connection to the fume extraction system
This chapter describes the eGROSS pro-x/eGROSS unit's connection to the fume extraction system.
The figure illustrates where to connect the Ø200mm (7.9in) pipe. Tighten the clamp. The image refers to the
left side part of the unit. Connect the opposite end to the laboratory's fume extraction system.
2.3.4. Connection to the fixative loading/unloading system
This chapter describes the eGROSS pro-x/eGROSS unit's connection to the fixative loading/unloading system.
On the eGROSS pro-x/eGROSS unit, it is possible to:
- Empty the fixative from the containers before cutting the samples.
- Load new fixative into the containers in which the baskets are positioned with the trays of the samples just
cut, in order to initiate the fixation process at room temperature inside the eGROSS pro-x/eGROSS unit.
The figure shows the connectors for the fixative loading/unloading containers when the unit is received.
Shorten the pipe to keep the expulsion line as linear as possible.
Avoid bending the pipe at excessive angles.
It is necessary to connect the unit to a fume extraction system.
The following are connected to aspiration: work surface, work surface's rear grille, grille
behind the wash basin, used and new fixative container cap and the bin (if present).