Application technology
43
Wash load...
...wide-neck
Wash load items with wide necks, e.g. beakers, wide neck Erlen-
meyer flasks and petri dishes, or cylindrical items, e.g. test tubes, can
be cleaned inside and out by rotating spray arms.
To do this the wash load is positioned in full, half or quarter inserts
and placed in an empty lower basket or an upper basket with a spray
arm.
...narrow-neck
Baskets with special injector modules are available for wash load
items with narrow necks, e.g. narrow neck Erlenmeyer flasks, round
bottomed flasks and measuring flasks.
The injector units and modules come with their own operating in-
structions.
When loading please note:
- Place petri dishes or similar items in the appropriate insert with the
dirty side facing towards the middle.
- Place pipettes with the pointed end facing downwards.
- Quarter segment inserts should be positioned at a minimum
1" (3 cm) distance from the edge of the upper or lower basket.
- Position quarter segment inserts for test tubes around the middle to
leave the corners of the upper or lower basket free.
- Use a cover net to avoid breakages if required.