6.
Attach the other end of the copper tubing to the gas fitting on the back of
the degasser.
Figure 1-3. Attaching Copper Tubing to the SmartPrep
7.
Open the gas bottle shut-off valve.
Use the regulator pressure control knob to set the pressure to 15 psi
(103 kPa).
8. Close the gas bottle shut-off valve. Check the pressure. If it decreases, gas is
probably leaking at the fitting. Tighten the fitting. Then check the pressure
again.
9.
Open the regulator shut-off valve. The pressure should stabilize after about
five minutes. If it does not, gas is probably leaking at the brass reducer fit-
ting or at the gas port on the rear panel of the degasser. Tighten the fittings,
then check the pressure again.
10. Open the gas bottle shut-off valve.
11. There are two RS-232 serial ports on the back panel of the degasser. They
are labeled with arrows indicating data flowing into the left connector and
data flowing out of the right connector.
Installation
SmartPrep
1-6
Sep 98
Summary of Contents for SmartPrep 065
Page 1: ...SmartPrep 065 Operator s Manual V2 02 065 42801 01 March 2002...
Page 4: ......
Page 8: ......
Page 10: ......
Page 26: ......
Page 28: ......
Page 60: ......
Page 61: ...CHAPTER 3 SMARTPREP MENUS File Menu SmartPrep Menu Reports Menu Options Menu...
Page 62: ......
Page 100: ......
Page 101: ...CHAPTER 4 MAINTENANCE...
Page 102: ......
Page 105: ...CHAPTER 5 ORDERING INFORMATION...
Page 106: ......
Page 108: ......
Page 109: ...APPENDIX A FORMS...
Page 110: ......
Page 112: ......
Page 114: ......
Page 115: ...APPENDIX B ERROR MESSAGES...
Page 116: ......
Page 127: ...INDEX...
Page 128: ......