Installation
Rear Panel Connections
1.
Insert one end of the power cord into the input power connector at the rear
of the degasser (refer to Figure 1-4) and the other end into an appropriate
power source.
2.
Attach an appropriate regulator to the gas bottle. Leave the gas bottle shut-
off valve closed until instructed otherwise.
3.
If the regulator has a 1/8-in. outlet, proceed to Step 4. If the regulator has a
1/4-in. outlet, attach the reducer fitting to the outlet of the regulator shut-off
valve.
Figure 1-2. Connecting Gas Bottle to the Regulator
4.
Tighten the regulator shut-off valve nut.
Do not over-tighten the fittings. Doing so can collapse the copper tubing and
cause a leak.
5.
Attach the copper tubing to the brass reducer fitting.
Pressure
Control
Knob
Bottle Shut-Off Valve
Two-Stage Pressure
Reducing Regulator
Brass Reducer
Fitting
Shut-Off
Valve Nut
Regulator Shut-Off
Valve
Copper Tubing
SmartPrep
Installation
Sep 98
1-5
Summary of Contents for SmartPrep 065
Page 1: ...SmartPrep 065 Operator s Manual V2 02 065 42801 01 March 2002...
Page 4: ......
Page 8: ......
Page 10: ......
Page 26: ......
Page 28: ......
Page 60: ......
Page 61: ...CHAPTER 3 SMARTPREP MENUS File Menu SmartPrep Menu Reports Menu Options Menu...
Page 62: ......
Page 100: ......
Page 101: ...CHAPTER 4 MAINTENANCE...
Page 102: ......
Page 105: ...CHAPTER 5 ORDERING INFORMATION...
Page 106: ......
Page 108: ......
Page 109: ...APPENDIX A FORMS...
Page 110: ......
Page 112: ......
Page 114: ......
Page 115: ...APPENDIX B ERROR MESSAGES...
Page 116: ......
Page 127: ...INDEX...
Page 128: ......