Page 40
Appendix | GenICam surfaceCONTROL Parameters | Version 1.5
surfaceCONTROL 3D SC2500 / SC2510
Name
Description
Documentation Text
EventErrorCode
Returns an error code for
the error(s) that hap-
pened.
This parameter returns the defect type for a measuring error. The
following error codes are possible:
- 1: Error Sensor Hardware: An error occurred in a hardware compo-
nent in the sensor. Contact Micro-Epsilon.
- 2: Error Sensor Acquisition: An error occurred during data acquisition
in the sensor. Contact Micro-Epsilon.
- 3: Error 3D Reconstruction: The 3D reconstruction could not be
calculated.
- 9: Error Critical Temperature: As temperatures are too high, further
measurements are prevented.
EventErrorMessage
Returns a detailed mes-
sage for the error.
In addition to the error type, an additional description of the measuring
error is returned if necessary.
User Set Control
UserSetSelector
Selects the feature User
Set to load, save or config-
ure. Default = factory
settings
Selects the user settings function for loading, saving or configuring.
Here, Default means default settings.
UserSetLoad[UserSetSelector]
Loads the User Set speci-
fied by UserSetSelector to
the device and makes it
active.
Loads the user setting specified by “”UserSetSelector”” into the sen-
sor and enables it.
UserSetSave[UserSetSelector]
Save the User Set speci-
fied by UserSetSelector to
the non-volatile memory of
the device.
Saves the user setting specified by “”UserSetSelector”” to the
sensor.
UserSetDefault
Selects the feature User
Set to load and make
active by default when the
device is reset.
Default = factory settings
Selects the user setting that is loaded and enabled by default
when the sensor is reset. Here, Default means factory settings.
DigitalIO Control
LineSelector
Selects the physical line
(or pin) of the external
device connector or the
virtual line of the Transport
Layer to configure.
Selection of the GPIO port to be configured.
LineMode[LineSelector]
Controls if the physical
Line is used to Input or
Output a signal.
Defines whether the selected GPIO port is an input or output.
LineInverter[LineSelector]
Controls the inversion of
the signal of the selected
input or output Line.
Enables inverting of the input or output signal for the selected
GPIO port.
LineStatus[LineSelector]
Returns the current status
of the selected input or
output Line.
Queries the status of the selected GPIO port.
LineStatusAll
Returns the current status
of all available Line signals
at time of polling in a sin-
gle bitfield.
Outputs the current status of all digital in-/outputs.
LineSource[LineSelector]
Selects which internal
acquisition or I/O source
signal to output on the
selected Line.
Selects the internal source for the output signal of the selected GPIO
port.
Summary of Contents for surfaceCONTROL SC2500