Page 27
optoCONTROL 1202
6.4
Limit Monitoring, Digital Outputs
OUT0
+U
B
0 V
OUT1
+U
B
0 V
Time [s]
OUT2
+U
B
0 V
If the receiver detects a signal below
the tolerance limit for example, the
according digital output is set. Also
LED‘s on the receiver indicate the
states of the digital outputs.
Thereto a measuring window must
be parameterized. The setup is ef-
fected
- at the receiver via the
“Teach“- key and the
, in the software tool
via the parameter
“Teach“ and
“TOLERANCE-VALUE“, see
software description.
Fig. 12 Switching behaviour of the digital outputs, output polarity (POLARITY) = DIRECT