BATTERIES
CHARGE
!
Attention:
Deep discharge very quickly leads to capacity loss. –
Therefore the batteries have to be charged at once
(1). Especially before a longer storage the batteries
are to be completely charged and because of the
constant own discharge to be recharged in-between.
Marginally increased transition resistance (> 0.2 Ohm)
in the charging cables and -plugs already prevent a
complete charging. Plugs that become hot during
charging are to be replaced.
BATTERY CHARGERS
☞
Note:
When changing to batteries with considerably dif-
ferent capacity also use a corresponding charger, so
that the charging periods remain limited and the
batteries are charged completely.
☞
Note:
When replacing the secondary plug (part-no. 206 917
100) a bridge (A), view inside view of the plug, must
be welded between the contacts 2 (earth) and 3.